RESTING
ECG
Microcomputer-controlled resting ECG.
Acquisition:
12 simultaneous leads, Standard or Cabrera
Pre-acquisition: Provides 10
seconds of instantaneous ECG acquisition
and printing in 3, 6 & 12 channel formats.
Signal
input: Type CF according to IEC
Sampling rate: 1000 Hz
Dynamic
range: ±10mV
Superimposed
DC voltage: (Polarization voltage) ±600mV
Paper Speed: 5/10/25/50 mm/s (direct)
Digital
Resolution:
5 µV
Frequency
response: 0.05 - 150 Hz
Common
mode rejection: > 140 dB
Input
impedance: > 100 MOhms
Leakage
current: < 10 µA
Pacemaker
detection sensitivity: All leads, duration 0.1-2.5 ms, amplitude>5 mV
ECG interpretation: Adult and Pediatric
SPIROMETRY
Predicteds:
Crapo, Morris, Knudson, Knudson76, Polgar, others. Extrapolated
predicted values.
Spirometry Measured Values:
FVC, FEV0.5, FEV1, FEV1/FVC, FEF25-75, PEF25%, FEF50%, FEF75%,
FIVC, FIV1, FIV1/FIVC, PIF, FIF50%, SVC, ERV, IRV, TV, MVV,
RR, TV. Pre vs. post medication comparison
Measurement Range:
Flow: 0 to +- 14 l/s
Volume: 0 to +1 11 l
Flow
Impedence: < 0.2 mbar*s/l at
12 l/s
Spirometry Standards:
P8000 Power meets or exceeds ATS, OSHA and NIOSH standards
|
SYSTEM
SPECIFICATIONS
Size:
39 x 33 x 10 cm / 15" X 13" X 3.9"
Weight: 5 Kg
(with paper) / 8.8 lbs
Power:
100-115V or 220-240V, 50/60Hz
Battery:
Built in 12V lead acid battery. ~ 4 hours of
use per charge.
Power
Consumption: Internal printing: 28VA max
Memory:
About 45 resting ECG or spirometry files
LCD Display:
Type: Backlit high resolution LCD
Size: 320 x 240 pixel, 3 1/2 x 5"
ECG: Selected lead group, operation
mode, lead check, heart
rate, AC filter, muscle filter, ADS, gain, speed, name,
ID,
3 ECG waveform
Spirometry:
Real time flow/volume loop or time/volume curve, parameters with
predicted, actual, and % difference values, interpretation, Up to 3
efforts, pre/post comparison
Monitor output:
Standard monitor interface
Internal printer:
High resolution digital printer
External printer:
Parallel port for connection to plain paper printers
Languages: English, Spanish,
French, others
|